True and False Statements
Custom Programs 7
37
Option Description
THEN Goto Next Phase Go to the next phase.
THEN Goto Phase 1
Go to
Phase
1
.
THEN Goto Phase 2
Go to
Phase 2
.
THEN Goto Phase 3
Go to
Phase 3
.
THEN Goto Phase 4 Go to Phase 4.
THEN Goto Phase 5
Go to
Phase 5
.
THEN Goto Done Go to Phase Done.
THEN REPEAT 0 times Repeat actions in the current phase by the number
of times specified. Available settings: 1-8 and
Forever
. If
Forever
is selected, the cycle repeats
for 255 times. Refer to IF Loop Count < Max and
the IF Loop Count > Max in
Test
.
THEN REST for 000:00:00
Apply no current for the specified time period (in
hhh:mm:ss format). The Resting xx:xx message
(code 19) is displayed.
After resting, the program goes to the next phase.
THEN FAIL Fail the battery. The battery is electrically
disconnected and the Fail 16 code will be
displayed with the FAIL light on.