4. If necessary, check the removed fuse with an ohmmeter to confirm diagnosis
and/or replace the station fuse.
5. Replace the upper cover of the C7000, (make sure the front edge of the top
cover is properly inserted into the notches in the plastic cups), and re-install the
screws to secure cover.
Extended Range Model (C7000ER)
1. Remove all batteries and adapters, turn off the analyzer and unplug power cord.
Wait 30 minutes to allow the voltage to discharge.
2. Unscrew the 2 screws located on the rear to remove the top cover. Disconnect the
wire on the power supply circuit board that is attached to the power assembly filter.
3. Remove the open fuse from the holder, taking care
not to touch any of the
electrical components
. The station fuses are located in the left rear corner of the
analyzer on the motherboard (refer to the diagram for standard model).
4. If necessary, check the removed fuse with an ohmmeter to confirm diagnosis
and/or replace the station fuse.
5. Reconnect the wire on the power supply board and re-install the screws to secure
cover.
70 C7000 & 7000ER Battery Analyzer V3.5