Detail Global Possible Reason(s) / Solution(s)
>2000 m
-
Resistance of battery is too high. Discharge battery before
performing the
OhmTest
.
A
Adapter data invalid FAIL 209 Invalid value in selected C-Code. Modify the C-Code or
delete and re-enter the C-Code.
The C-Code is not compatible with the firmware; the
adapter could have been configured with a newer firmware.
Delete and re-enter the C-Code, reset the Extended
C-Code to default values or upgrade the analyzer with the
latest firmware.
Adapter inserted ADAPT IN Adapter is detected on a station. Global message should
be displayed. If not, remove and re-insert adapter.
If the message persists, re-enter the C-Code.
Adapter not set up FAIL 208 Adapter is not programmed with a C-Code or the C-Code
is invalid. Re-enter a C-Code.
Adapter removed ADAPT OUT Adapter is removed from a station. If message persists,
press
Esc
.
B
Bad adapter FAIL 210 Adapter is not recognized by the analyzer. Remove and
re-insert the adapter. Clean the gold contacts on the
underside of the adapter.
Reset the Extended C-Code to default values or delete and
re-enter a C-Code.
Bad fuse or driver FAIL 160 The analyzer is not able to supply current to the battery
through its current drivers. Very low values of current (about
0mAh) may be displayed on the analyzer before this code
appears and processing stops. Check and replace the
station fuse. If fuses are ok, then the current drivers could
have failed on the analyzer. Before calling Cadex for
service, check the following items for other causes of
faults:
Ensure that battery and adapter contacts are clean.
Verify that mAh rating of the battery matches the C-Code
mAh setting.
The battery may not be able to accept the current load.
Lower or raise the Charge rate in the Extended C-Code.
The battery may be intrinsically safe. Use I/S settings.
Refer to Intrinsically Safe Settings.
Try a different battery in the same station.
Ω