Getting Help/Problem Solving 15
77
Problem Possible Reason(s) / Solution(s)
9. Program did not resume
after a power failure.
Check if an option board (OP-70 or OP-72) is
installed. The option board sticker is at the
back of the analyzer. The option board
configuration is also displayed during start-up.
The analyzer was saving the battery information
when the power failed. Battery information is
saved once a minute. If the power fails during
the information save, the program will not
continue when power resumes.
Replace the backup battery.
10. Single cell gives inconsistent
results.
Calibrate the station.
Set the charge method in the Extended
C-Code to DC Charge due to low terminal
voltage.
Single cells are discharging and/or charging
too quickly. Lower the Discharge rate and
Charge rate to about 0.2C.
11. Capacity readings between
stations are inconsistent.
Verify that the C-Codes are the same between
stations.
Calibrate the stations.
12. Resistance readings are
consistently high.
Calibrate the station.
The battery or adapter contacts may be
damaged.
The battery is faulty.
13. Battery passes on analyzer
but fails in the field.
Use the
OhmTest
. If the resistance is high, the
battery may not be able to hold charge.
Use the Self Discharge program to check the
difference of the last two readings. If it is more
than 30%, the battery fails.