Chapter 12 System Calibration
There are two kinds of calibration that should be performed on the Cadex analyzer
on a regular basis: voltage calibration and adapter calibration.
Voltage Calibration
Voltage calibration should be performed yearly and a calibration kit is required.
The C7000-C Series of Battery Analyzers uses Calibration Kit P/N 92-770-
0212 (See Appendix D - Parts and Accessories, page 154).
To calibrate the voltage
1. Remove all Battery Adapters.
2. From the global display, press MENU.
3. Press or to scroll to SYSTEM SETUP and then press ENTER.
4. Press or to scroll to CALIBRATE VOLTAGE and then press ENTER.
5. Firmly insert the four voltage calibration adapters, one into each
station. (If calibrating a C7200, only two voltage calibration adapters
are necessary)
6. Wait for the prompt, and then remove the calibration adapters.
7. Press MENU to return to the global display.
8. Press FN, and then press 7. The “#” sign appears in the upper right
corner of the startup screen, indicating successful completion of
voltage calibration.
Cadex Electronics Inc. C7000-C Series Battery Analyzers .
112