STATION saves any changes made to the program settings or the target capacity to
the battery station. These changes override any settings in the active C-code.
5. Press ENTER.
6. Press or to scroll to YES to accept changes and press ENTER.
7. Press ESC twice to return to the global display.
Note To use all of the settings from the active C-code, this option must be set to
C-CODE. Setting it to C-CODE will erase all settings saved to the stations.
Resuming Service After Power Failure
If there is a power failure, the back-up memory in the analyzer keeps service data
for one hour. If the power failure lasts less than an hour, battery service resumes
from where it left off. If the power failure lasts longer than one hour, the battery
resumes from the beginning. This time can be modified:
To change the Power Failure Resume Hours
1. From the global display, press MENU.
2. Press or to scroll to SYSTEM SETUP, press ENTER.
3. Press or to scroll to POWER FAILURE RESUME HOURS, press ENTER.
4. Press or to scroll to the required setting:
5. Press ENTER. Press ENTER again to save the settings.
6. Press ESC twice to return to the global display.
81