Parameter Default Setting
Max. Charge Voltage
3.60V/cell
End of Charge
0.03 C
End-Of-Discharge
2.00V/cell
Extended C-code Settings for Special Cases
This section lists recommended parameter settings for specific types of batteries,
where the recommended settings are different from the default settings.
Intrinsically Safe Batteries
Before servicing intrinsically safe (I/S) batteries, set the following C-code
parameters as specified:
Parameter Setting
Charge
0.10 C*
Discharge
0.10 C*
Charge Method
DC charge
End-Of-Discharge
0.96V/cell
Hawker SLA Batteries
Before servicing Hawker SLA batteries, set the following C-code parameters as
specified below (any parameters not listed here should be left at their default
settings):
Parameter Setting
Discharge
0.40 C
* Recommended settings. You can try setting the charge and discharge rates to higher values to speed up
servicing; however, if fault codes appear, you must lower the rates.
Cadex Electronics Inc. C7000-C Series Battery Analyzers .