Ready—No Charge Ends the program. The green READY light is on, and
no trickle charge is applied.
Ready—Trickle Charge Trickle-charge the battery using the trickle-charge
rate in the extended C-code. Includes the standby
maintenance cycle, which performs a discharge/
charge every 30 days (180 days for SLA).
Trickle Charge Trickle-charge the battery using the trickle-charge
rate in the extended C-code.
Creating and Editing Custom Programs
To create or edit a custom program
1. From the global display, press MENU.
2. Press or to scroll to PROGRAM SETTINGS and then press ENTER.
3. Press or to scroll to CUSTOM PROGRAMS and then press ENTER.
4. Press or to scroll to the program to create or edit, for example,
CUSTOM 1.
5. Press EDIT. The cursor appears on Phase 1.
6. Press to move the cursor to the Cycle 1 field.
7. Program the action to be completed during this cycle:
Press or to select the action.
Press .
If necessary, use the direction keys or the number pad to enter any additional
settings required for the cycle and then press .
1. Repeat step 7 for Cycle 2 of the phase.
2. Program the test for the phase:
Press or to select the IF statement, the test to be performed. (see Test, page
102).
Press .
105