Parameter Setting
Max. Standby Voltage
2.35V/cell
Max. Charge Voltage
2.60V/cell
End-Of-Discharge
1.62V/cell
Gell SLA Batteries
Use the SLA default settings for Gell batteries.
Managing C-codes
If the ASK FOR BATTERY C-CODE option is set to YES, the desired C-code must be
selected and the appropriate parameters entered before a battery is inserted or
before battery processing begins (see To set the user input required by the
Cadex analyzer before battery service begins, page 77).
Select a C-code
1. From the global display, press the station key.
2. Press or to scroll to the C-code to select.
3. Press ENTER.
4. Press or to scroll to YES to accept changes and press ENTER.
5. Press ESC twice to return to the global display.
The “*” appears beside the C-code number on the display to indicate that the C-
code is active.
Display a C-code
1. From the global display, press the station key.
2. Use the or key to scroll to the C-code to display. Basic C-code
parameters appear on the same screen.
61