7. Press ENTER.
8. Press or to scroll to YES to accept changes and press ENTER.
9. Press ESC twice to return to the global display.
Runtime
Function: The Runtime program simulates field application with three adjustable
load currents and time intervals.
Sequence: The battery is charged. The program then repeatedly runs three
different loads (see instructions below to set these loads) until the battery is fully
discharged (i.e. The end-of-discharge setting in the C-code is reached).
Results: The detailed display shows the battery run-time (hours and minutes). It
also shows battery or cell voltage; analyzer charge or discharge current (in mA);
OhmTest results; battery temperature and duration of service.
Duration: Approximately 8 – 10 hours for a NiCd battery using default C-codes.
To edit the Runtime settings
1. From the global display, press MENU.
2. Press or to scroll to PROGRAM SETTINGS and then press ENTER.
3. Press or to scroll to RUNTIME SETTINGS and then press ENTER.
4. The cursor is at Discharge 1 (the first load).
5. Press or or use the number pad to set a load between 1 and
100 for the Discharge %. This is a percentage of the Discharge Rate in
the C-code. The default is 100.
6. Press to move to Minutes – the time this load is applied. Press
or or use the number pad to set the time between 1 and 59
minutes. The default is 1.
7. Press to move to the load setting for the Discharge 2 (second
phase). Repeat steps 4 and 5.
37