16
3. Specifications
Series name
CAUW-D series
(Dual range type)
CAUW series CAUX Series CAUY series
Model name
CAUW220D CAUW120D
AUW320
AUW220
AUW120
AUX 320
AUX220
AUX120 CAUY220
AUY120
Capacity
220g/82g 120g/42g 320g 220g 120g 320g 220g 120g 220g 120g
Minimum display
.1mg/0.01mg
.1mg/0.01mg 0.1mg 0.1mg 0.1mg 0.1mg 0.1mg 0.1mg 0.1mg 0.1mg
Repeatability
(standard deviationσ)
≤0.1mg
(220g range)
≤0.05mg
(82g range)
≤0.1mg
(120g range)
≤0.02mg
(42g range)
≤0.15mg ≤0.1mg ≤0.1mg ≤0.15mg ≤0.1mg ≤0.1mg ≤0.1mg ≤0.1mg
Linearity
±0.2mg
(220g range)
±0.1mg
(82g range)
±0.2mg
(120g range)
±0.03mg
(42g range)
±0.3mg ±0.2mg ±0.2mg ±0.3mg ±0.2mg ±0.2mg ±0.2mg ±0.2mg
Response time
stabilization time, typical)
3s
(220g range),
15s
(82g range)
3s
(120g range),
12s
(42g range)
3 s
Calibration weight Built-in None
Operation tem-
perature range
5 to 40 deg.C
Temperature
coefficient of sensitivity
(10 to 30 deg.c)
±2ppm/deg.C (when PSC function is Off) ±2ppm/deg.C
Sensitivity stability
(10 to 30 deg.c)
±2ppm (when PSC function is On)
Pan size 80mm dia Approx.
Main body dimensions 220mm W X 330mm D X 310mm H
Main body Weight 7kg Approx.
Power requirements DC, 10 to 15.5V, 600mA (Plug polarity: center negative)
Data I/O RS-232C
Display backlight
y y y
PSC automatic
span calibration
y y y y y y y y
Clock-CAL
y y y y y
Built-in clock
y y y y y y y y
GLP/GMP/ISO
calibration report
y y y y y y y y
WindowsDirect
y y y y y y y y y y
Interval timer
output
y y y y y y y y
RS-232 C I/F
y y y y y y y y y y
Specific gravity
measurement
software, piece
ounting, % display,
unit conversion
y y y y y y y y y y
Analog display
y y y y y y y y y y
F
u
n
c
t
i
o
n
Battery (option)
y y y y y y y y y y