63
Date output, ID number etc. of electronic printer EP-50 or EP-90 should be
Turned off when calibration report is produced.
10.4.2 Balance ID Setting
This setting is for the balance ID number that is output along with the calibration report.
1 From mass display, press the [CAL] key
repeatedly until “SEttinG” appears.
Press the [O/T] key. “CAL dEF” appears.
2 Press the [CAL] key repeatedly until
“id: ****” appears. (The **** is a number.)
Press the [O/T] key. In the upper part of the
display panel, the MENU symbol and the #
symbol appear in order to indicate numerical
input mode. The leftmost digit of **** blinks.
The numeral of the blinking digit can be
changed.
3 Press the [UNIT] key to increase the value of
the blinking digit by 1. Press the [PRINT] key
to set that digit and move the blinking digit
one place to the right. When the desired setting
is entered, press the [O/T] key to confirm the
balance ID number.
4 Press the [POWER] key to return to
“id: ****”. (No blinking)
5 Press the [POWER] key again to return to
the mass display.
(Example)
(Example)