60
10.3.5 Inputting External Calibration Weight Value for E-CAL
The exact value (conventional mass value) of the operator’s calibration weight to be used
for E-CAL and E-tESt procedures can be entered.
1 From mass display, press the [CAL] key
repeatedly until “SEttinG” appears.
Press the [O/T] key. “CAL dEF” appears.
2 Press the [CAL] key repeatedly until “CAL
SEt” appears and press the [O/T] key. One of
these numbers will appear, depending on the
model: 50.0000g, 100.0000g, 200.0000g, or
300.0000g. In the upper part of the display
panel, the MENU symbol and the # symbol
appear in order to indicate numerical input
mode. The leftmost numeral blinks.
The blinking digit may be changed.
3 When the [UNIT] key is pressed, the value of
the blinking digit increases by 1 at a time. Press
the [PRINT] key to move the blinking digit one
place to the right. Enter the desired value.
4 Press the [O/T] key to set the external
calibration weight value. The display proceeds
to “SEt” and then to “CAL SEt”.
5 Press the [POWER] key again to return to
the mass display.
Calibration Weight Value Range
220D 75g to capacity
CAUW-D series
120D 35g to capacity
320 145g to capacity
220 95g to capacity
Other series
120 45g to capacity
In the CAUW-D series the conventional mass value may be input for up to
five decimal places.
(Example)
(Example)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CAS CAU Series and is the answer not in the manual?

CAS CAU Series Specifications

General IconGeneral
BrandCAS
ModelCAU Series
CategoryScales
LanguageEnglish

Related product manuals