13.5 Interval Timer (CAUW-D/CAUW/CAUX series only)
This function automatically outputs the measurement values of the balance at set time
intervals.
1 Press the [CAL] key repeatedly from the
mass display until “FUnC.SEL” appears. Press
the [O/T] key to display “CAL”.
2 Press the [CAL] key repeatedly until “int”
appears.
3 Press the [O/T] key. “int: **:**” appears.
(**** are numbers)
The MENU symbol and the # symbol appear
in order to show number entry mode. **:**
shows the current interval setting as [MM:SS]
(where MM is minutes and SS is seconds). The
leftmost digit blinks.
4 Pressing the [UNIT] key increases the blinking
digit by 1. Pressing the [PRINT] key moves the
blinking digit one place to the right. Set minutes
up to 99 and seconds up to 59.
5 Press [O/T] key to set.
6 Press the [POWER] key repeatedly to return to
the mass display. The letter “T” and STAND-BY
symbol appear indicating interval output standby.
7 Press the [PRINT] key. Automatic output
commences and continues at the set interval.
8 To pause, press the [POWER] key.
9 To discontinue the Interval Timer printing
function, hold down the [POWER] key during
mass display for about three seconds until
“APL End” appears. The letter “T” and STAND-
BY symbol will disappear.
(Example)
(Example)
(Example)
(Example)
(When set to 1 minute, 30 seconds)