51
10.2.2 Span Calibration With External Weights
“E-CAL” (The balance will be adjusted with external calibration weights)
1 From the mass display, press the [CAL]
key once. If the display shows “E CAL”
(Preset calibration method is “E-CAL”),
jump to Step 4.
If the display shows another item
(Preset calibration method is not “E-CAL”),
go on to Step 2.
(Refer to 10.3.1 for preset calibration method)
2 Press the [CAL] key repeatedly until
“FUnC.SEL” appears, then press the [O/T]
key to display “CAL”.
3 Press the [O/T] key. The display shows
“E CAL”.
4 While “E-CAL” is shown, press the [O/T] key.
Span calibration begins and the zero display
blinks.
5 Recheck that there is nothing on the pan and
wait.
6 The set weight value (Refer to 10.3.5) appears,
blinking.
7 Place the weight indicated on the pan. Wait
until the blinking zero display reappears.
(It may take about 30 seconds.)
8 Remove the calibration weight.
After “CAL End” appears for several seconds,
the display returns to the mass display and the
span calibration is complete.
O
/
T
1
d
/
1
0
d
U
N
I
T
P
R
I
N
T
P
O
W
E
R
BR
K
C
A
L
M
E
N
U
g
(Example)