81
1 Press the [CAL] key repeatedly from the mass
display until “FUnC.SEL” appears. Press the
[O/T] key. The display will show “CAL”.
2 Press the [CAL] key repeatedly until “Addon:
**” appears. The “**” shows the current setting,
“on” for on, “oF” for off.
3 Press the [O/T] key. The display shows “Adon-
on” and after this, pressing the [CAL] key
switches the displays between “Adon-on” and
“Adon-oF”. When the current setting is
displayed, the stability mark (¨) appears.
Press the [O/T] key when the desired setting is
shown, to change the setting.
4 Press the [POWER] key to return to“Addon:**”
5 Pressing the [POWER] key again returns to the
mass display.
6 When the Add-on Mode is set up, the add-on
mode symbol appears in the mass display.
x No more than one of the four application modes, Auto Print (13.4), Interval
Timer (13.5), Addon Mode (13.6), Formulation Mode (13.7), can be set ON
at the same time.
x When Add-on mode is in use, fully-automatic span calibration by PSC (10.3.2)
or Clock-CAL (10.3.3) is not performed. Instead, Weight symbol ( )
keeps blinking when span calibration is necessary. Span calibration can be
performed between sets of measurements.