57
GB
INSTALLATION
RX DC X-RAY UNIT
During the update the error code E108 will appear on the handheld device. At the end, simply press one key on
the handheld device to eliminate the error.
11. The device will then automatically start the rmware update procedure. Turn the handheld unit on: the display
shows the progress status (see gure) and the machine emits a beep sound every 2".
12. Once the procedure is completed, the device will be automatically restarted (approximately 6 minutes are
required in wireless mode and 16 minutes in wired mode).
13. Check on P03 that the rmware versions of the handheld unit, X-ray head and basic board are aligned and
correspond to those of the selected update package.