GPIB/RS232 Interface (IEEE-488.2)
6-13
[:SOURce]:SAFety[:CHANnel]<m>:RESult:ALL:TIME[:ELAPsed]:DWELl?
It queries the DWELL time of all STEPs for test when the Screen sets to On. The format
of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:DWEL?
The tester returns 5.000000E+00, 5.000000E+00, 5.000000E+00,
1.000000E+00, 1.000000E+00, 1.000000E+00, 1.000000E+00,
2.000000E+00, 2.000000E+00, 2.000000E+00.
Description: The returned string is the dwell time tested of Address 0 host for CH5
STEP1~STEP10, that is 5.0sec, 5.0sec, 5.0sec, 1.0sec,1.0sec,
1.0sec, 1.0sec, 2.0sec, 2.0sec, 2.0sec.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:ALL:TIME[:ELAPsed]:FALL?
It queries the time of voltage fall for all STEPs when the Screen sets to On. The format
of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:FALL?
The tester returns 1.500000E+00, 1.500000E+00, 1.500000E+00,
2.000000E+00, 2.000000E+00, 2.000000E+00, 5.000000E-01,
5.000000E-01, 5.000000E-01, 5.000000E-01.
Description: The returned string is the voltage fall time tested of Address 0 host for
CH5 STEP1~STEP10, that is 1.5sec, 1.5sec, 1.5sec, 2.0sec,2.0sec,
2.0sec, 0.5sec, 0.5sec, 0.5sec, 0.5sec.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:ALL:TIME[:ELAPsed]:RAMP?
It queries the time required for voltage to ramp up of all STEPs when the Screen sets to
On. The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is
Frame index in the range of 0~9 and the last two digits are Channel index in the range of
01~10 (19020/19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:RAMP?
The tester returns 5.000000E+00, 5.000000E+00, 5.000000E+00,
1.000000E+00, 1.000000E+00, 1.000000E+00, 2.000000E+00,
2.000000E+00, 2.000000E+00, 2.000000E+00
Description: The returned string is the time required for voltage to ramp up to the
set in CH5 STEP1~STEP10 on Address 0 host, that is 0.5sec, 0.5sec,
0.5sec, 1.0sec,1.0sec, 1.0sec, 2.0sec, 2.0sec, 2.0sec, 2.0sec.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:ALL:TIME[:ELAPsed] [:TEST]?
It queries the test time of all STEPs when the Screen sets to On. The format of variable
<m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the range of 0~9
and the last two digits are Channel index in the range of 01~10 (19020/19021/ 19022) or
01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is 001 to indicate
Channel 1 of the Master.