Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
Users Manual
6-14
Example: Input the command SAF:CHAN005:RES:ALL:TIME?
The tester returns 3.000000E+00, 3.000000E+00, 1.000000E+00,
2.500000E+00, 2.500000E+00, 1.000000E+00, 1.000000E+00,
0.500000E+00, 3.000000E+00, 1.000000E+00.
Description: The returned string is the time required for testing for CH5
STEP1~STEP10 of Address 0 host, that is 3.0sec, 3.0sec, 1.0sec,
2.5sec,2.5sec, 1.0sec, 1.0sec, 0.5sec, 3.0sec, 1.0sec.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:AREPort < boolean > | ON | OFF
It sets if reporting the test result automatically (only valid for RS232 interface.) This
command is only valid for this Tester, so the Address 0 host <m> can use 001, Address
1 host <m> can use 101 and so forth. The Address 9 host <m> can use 901. The
returned data is Total Pass/Fail state and its format is string in “PASS” or “FAIL”.
Example: Input the command SAF001:RES:AREP ON
Description: It means to set the Address 0 host to return the test result
automatically after the test is done. If the test result of all channels of
Address 0 host is PASS, it returns “PASS” orFAIL” will be returned.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:AREPort?
It queries the device if reporting the test result automatically. The return is 1 or 0.
(Applicable for RS232 interface only.)
Example: Input the command SAF001:RES:AREP?
The tester returns 1.
Description: The returned 1 means Address 0 host will report the test result
automatically once the test is done.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:COMPleted?
It queries the device if all settings are executed completely. The return is 1 or 0.
Example: Input the command SAF001:RES:COMP?
The tester returns 1.
Description: The returned 1 means Address 0 host has completed the execution of
all settings.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:STEP<n>[:JUDGment]?
It queries the judgment of specified STEP when the Screen sets to On. The format of
variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master. Return format: First Step Result, Second Step
Result, ..., Last Step Result. The table below lists the definition of Code:
Code of Test Result
Mode
AC
DC
IR
OSC
ALL
Code
HEX
DEC
HEX
DEC
HEX
DEC
HEX
DEC
HEX
DEC
STOP
70
112
TESTING
73
115
PASS
74
116
HIGH FAIL
21
33
31
49
41
65
LOW FAIL
22
34
32
50
42
66
ARC FAIL
23
35
33
51