GPIB/RS232 Interface (IEEE-488.2)
6-15
OCP
24
36
34
52
44
68
64
100
SHORT FAIL
61
97
OPEN FAIL
62
98
Example: Input the command SAF:CHAN003:RES:STEP3?
The tester returns 116.
Description: The returned result indicates the judgment of CH3 STEP 3 for
Address 0 host is PASS.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:STEP<n>:MMETerage?
It queries the MEASURE METER reading of specified STEP when the Screen sets to
On. The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is
Frame index in the range of 0~9 and the last two digits are Channel index in the range of
01~10 (19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN004:RES:STEP3:MMET?
The tester returns 4.000000E-05.
Description: The returned result indicates the MEASURE METER reading of CH4
STEP 3 for Address 0 host is 0.04mA.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:STEP<n>:OMETerage?
It queries the OUTPUT METER reading of specified STEP when the Screen sets to On.
The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame
index in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN001:RES:STEP3:OMET?
The tester returns 1.000000E+02.
Description: The returned result indicates the OUTPUT METER reading of CH1
STEP 3 for Address 0 host is 0.100kV.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:STEP<n>:TIME[:ELAPsed]: DWELl?
It queries the DWELL time tested by specified STEP when the Screen sets to On. The
format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index
in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP1:TIME:DWEL?
The tester returns 5.000000E-01
Description: The returned result indicates the DWELL tested by CH5 STEP 1 for
Address 0 host is 0.5sec.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:STEP<n>:TIME[:ELAPsed]: FALL?
It queries the time of voltage fall in specified STEP when the Screen sets to On. The
format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index
in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP4:TIME:FALL?