EasyManua.ls Logo

Chroma 19053 - Page 81

Default Icon
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPIB Operation Description (Option)
: OPEN?
: SOURce: SAFEty: STEP: OSC: LIMit: OPEN?
This command queries selected STEP which setting percentage is
judged by open circuit as detecting short/open circuit.
Example: Input command “SAFE: STEP 4: OSC: LIM: OPEN?”
Return message “3.000000E-01”
Description: Return message “3.000000E-01” means open circuit
judgment percentage of the main unit STEP 4 as
detecting short/open circuit is 30%.
:SHORt <numeric_value>
: SOURce: SAFEty: STEP: OSC: LIMit: SHORt
This command sets selected STEP which setting percentage is
judged by short circuit as detecting short/open circuit.
Example: Input command “SAFE: STEP 4: OSC: LIM: SHOR 3”
Description: This command means short circuit judgment
percentage of the main unit STEP 4 as detecting
short/open circuit is set to 300%.
SHORt?
: SOURce: SAFEty: STEP: OSC: LIMit: SHORt?
This command queries selected STEP which setting percentage is
judged by short circuit as detecting short/open circuit.
Example: Input command “SAFE: STEP 4: OSC: LIM: SHORt?”
Return message “3.000000E+00”
Description: Return message “3.000000E+00” means short circuit
judgment percentage of the main unit STEP 4 as
detecting short/open circuit is 300%.
: CHANnel
[: HIGH] <channel_list>
: SOURce: SAFEty: STEP: OSC: CHANnel: HIGH
This command sets the status of scanning test high voltage output
channel.
Example: Input command “SAFE: STEP 4: OSC: CHAN(@(1,3))”
Description: This command means output channel of scanning test
of the main unit STEP 4 is set to channel 1 and 3 HIGH
output.
Example: Input command “SAFE: STEP 4: OSC: CHAN(@(0))”
Description: This command means output channel of scanning test
of the main unit STEP 4 is set to OFF.
[: HIGH]?
: SOURce: SAFEty: STEP: OSC: CHANnel: HIGH?
This command queries the status of scanning test high voltage
output channel.
Example: Input command “SAFE: STEP 4: OSC: CHAN?”
Return message “(@(1,3))”
Description: Return message “(@(1,3))” means output channel
status of scanning test of the main unit STEP 4 is
channel 1 and 3 HIGH output.