HIPOT Tester 19051/19052/19053/19054 User’s Manual
: LOW <channel_list>
: SOURce: SAFEty: STEP: OSC: CHANnel: LOW
This command sets the status of scanning common test channel
(RTN/LOW) output.
Example: Input command “SAFE: STEP 4: OSC: CHAN: LOW
(@(2,4))”
Description: This command means output channel of scanning test
of the main unit STEP 4 is set to channel 2 and 4 LOW
output.
Example: Input command “SAFE:STEP 4:DC:CHAN:LOW (@(0))”
Description: This command means output channel of scanning test
of the main unit STEP 4 is set to OFF.
: LOW?
: SOURce: SAFEty: STEP: OSC: CHANnel: LOW?
This command queries the status of scanning common test channel
(RTN/LOW) output.
Example: Input command “SAFE: STEP 4: OSC: CHAN: LOW?”
Return message “(@(2,4))”
Description: Return message “(@(2,4))” means output channel
status of scanning test of the main unit STEP 4 is
channel 2 and 4 LOW output.
: PAuse
: [:MESSage] <string>
: SOURce:SAFEty:STEP:PAuse:MESSage
This command sets the string of message.
Example: Input command “SAFE: STEP 5: PA: MESS CHROMA”
Description: This command means message string of the main unit
STEP 5 is set to CHROMA.
: [:MESSage] ?
: SOURce:SAFEty:STEP:PAuse:MESSage?
This command queries the setting string of message.
Example: Input command “SAFE: STEP 5: PA: MESS?”
Return message “CHROMA”
Description: Return message “CHROMA” means message string of
the main unit STEP 5 is “CHROMA”.
: UTSIgnal < boolean > | ON | OFF >
: SOURce:SAFEty:STEP:PAuse:UTSIgnal
This command sets the status of UNDER TEST SIGNAL.
Parameter is ON or 1 indicates the setting ON.
Parameter is OFF or 0 indicates the setting OFF.
Example: Input command “SAFE: STEP 5: PA: UTSI ON”
Description: This command means the status of UNDER TEST
SIGNAL of the main unit STEP 5 is set to ON.
: UTSIgnal?
: SOURce:SAFEty:STEP:PAuse:UTSIgnal?
This command queries the status of UNDER TEST SIGNAL.
Return 1 indicates the setting ON.
Return 0 indicates the setting OFF.
5-32