GPIB Operation Description (Option)
Example: Input command SAFE: STEP 5: PA: UTSI ON
Return message “1”
Description: Return message “1” means the status of UNDER
TEST SIGNAL of the main unit STEP 5 is ON.
: TIME
[:TEST] <numeric_value>
: SOURce: SAFEty: STEP: PAuse: TIME: TEST
This command sets selected STEP which test needed time.
Example: Input command “SAFE: STEP 5: PA:TIME 5”
Description: This command means test needed time of the main unit
STEP 5 is set to 5.0 sec.
[:TEST]?
: SOURce:SAFEty:STEP :PA:TIME:TEST?
This command queries selected STEP which test needed time.
Example: Input command “SAFE: STEP 5: PA:TIME?”
Return message “5.000000E+00”
Description: Return message “5.000000E+00” means test needed
time of the main unit STEP 5 is 5.0 sec.
: PRESet
: TIME
: PASS <numeric_value>
: SOURce: SAFEty: PRESet: TIME: PASS
This command sets the buzzer sound continuous time when pass.
<numeric_value> is 0.2-99.9(included).
Example: Input command “SAFE: PRES: TIME: PASS 1”
Description: This command means buzzer sound continuous time
when pass is set to 1 sec.
: PASS?
: SOURce: SAFEty: PRESet: TIME: PASS?
This command queries buzzer sound output continuous time when
pass.
Return value is 0.2-99.9(included).
Example: Input command “SAFE: PRES: TIME: PASS?”
Return message “1.000000E+00”
Description: Return message “1.000000E+00” means buzzer sound
continuous time when pass is 1 sec.
: STEP <numeric_value>|KEY
: SOURce: SAFEty: PRESet: TIME: STEP
This command sets the interval time between STEP and STEP.
<numeric_value> is a value or character KEY from 0.1 to 99.9
(included).
Example: Input command “SAFE: PRES: TIME: STEP 0.5”
Description: This command means the interval time between STEP
and STEP is set to 1 sec.
: STEP?
: SOURce: SAFEty: PRESet: TIME: STEP?
This command queries the interval time between STEP and STEP,