HIPOT Tester 19051/19052/19053/19054 User’s Manual
or the next start command to execute the next STEP.
Return value is 0.1-99.9 (included).
Example: Input command “SAFE: PRES: TIME: STEP?”
Return message “5.000000E-01”
Description: Return message “5.000000E-01” means the interval
time between STEP and STEP is 0.5 sec.
: RJUDgment <Boolean> |ON|OFF
: SOURce: SAFEty: PRESet: RJUDgment
This command sets the status of RAMP JUDGMENT.
Parameter is ON or 1 indicates the setting ON.
Parameter is OFF or 0 indicates the setting OFF.
Example: Input command “SAFE:PRES:RJUD ON”
Description: This command means the status of RAMP JUDGMENT is
set to ON.
: RJUDgment?
: SOURce: SAFEty: PRESet: RJUDgment?
This command queries the status of RJUDGMENT.
Return 1 indicates the setting ON.
Return 0 indicates the setting OFF.
Example: Input command “SAFE: PRES: RJUD?”
Return message “1”
Description: This command means the queried setting result of RAMP
JUDGMENT is ON.
: AC
: FREQuency <numeric_value>
: SOURce: SAFEty: PRESet: AC: FREQuency
This command sets the output voltage frequency when testing AC
withstand voltage.
<numeric_value> is the value of 50 or 60.
Example: Input command “SAFE: PRES: AC: FREQ 60”
Description: This command means the output voltage frequency when
testing AC withstand voltage is set to 60Hz.
: FREQuency?
: SOURce: SAFEty: PRESet: AC: FREQuency?
This command queries the output voltage frequency when testing AC
withstand voltage.
Return value is 50 or 60.
Example: Input command “SAFE: PRES: AC: FREQ?”
Return message “6.000000E+01”
Description: Return message “6.000000E+01” means the output voltage
frequency when testing AC withstand voltage is 60Hz.
: WRANge
[: AUTO] <boolean>|ON|OFF
: SOURce: SAFEty: PRESet: WRANge: AUTO
This command sets withstand voltage auto range function is ON or OFF.
Example: Input command “SAFE: PRES: WRAN OFF”
Description: This command means withstand voltage auto range function
is set to OFF.
5-34