GPIB Operation Description (Option)
[: AUTO]?
: SOURce: SAFEty: PRESet: WRANge: AUTO?
This command queries withstand voltage auto range function is ON or
OFF.
Return 1 or 0 (0 means withstand voltage auto range function is OFF, 1
means withstand voltage auto range function is ON).
Example: Input command “SAFE: PRES: WRNG?”
Return message “0”
Description: Return message “0” means withstand voltage auto range
function is OFF.
: AGC
[: SOFTware] <boolean>|ON|OFF
: SOURce: SAFEty: PRESet: AGC: SOFTware
This command sets software AGC is ON or OFF.
Example: Input command “SAFE: PRES: AGC ON”
Description: This command means software AGC function is set to ON.
[: SOFTware]?
: SOURce: SAFEty: PRESet: AGC: SOFTware?
This command queries software AGC is ON or OFF.
Return 1 or 0 (0 means AGC is OFF, 1 means AGC is ON).
Example: Input command “SAFE: PRES: AGC?”
Return message “1”
Description: Return message “1” means software AGC function is ON.
: GCONtinuity <numeric_value>|ON|OFF
: SOURce: SAFEty: PRESet: GCONtinuity
This command sets GR CONT. parameter of PRESET.
Parameter 0 or OFF indicates OFF, parameter 0.1 or ON indicates KEY and
parameter 0.2-99.9 sec indicates start.
Note
“start” is the count down second you need under the stage of
detecting DUT connection.
Example: Input command “SAFE: PRES: GCON ON”
Description: This command means GR CONT. function is set to ON.
: GCONtinuity?
: SOURce: SAFEty: PRESet: GCONtinuity?
This command queries if grounding resistance test ON or OFF, or enters the
status of SMART START.
Return OFF or ON or setting time (OFF indicates grounding resistance don’t
open, ON indicates grounding resistance open. When the time is responded,
it indicates to enter the status of SMART START.
Example: Input command “SAFE: PRES: GCON?”
Return message “1”
Description: Return message “1” means GR CONT. function is ON.
: GFI
[: SWITCH] <Boolean>|ON|OFF
: SOURce: SAFEty: PRESet: GFI: SWITCH
This command sets GFI switch is ON or OFF.
Example: Input command “SAFE: PRES: GFI OFF”
5-35