Remote Operation
4-17
Query Example: CURR:DYN:L1?
CURR:DYN:L1? MAX
CURR:DYN:L1? MIN
CURRent:DYNamic:L2
Type: Channel-Specific
Description: Set the load current during T2 period for constant current dynamic
mode.
Setting Syntax: CURRent:DYNamic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:DYN:L2 20 Set the dynamic load parameter
L2 = 20A.
CURR:DYN:L2 10A Set the dynamic load parameter
L2 = 10A.
CURR:DYN:L2 MAX Set the dynamic load parameter
L2 = maximum value.
CURR:DYN:L2 MIN Set the dynamic load parameter
L2 = minimum value.
Query Syntax: CURRent:DYNamic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: CURR:DYN:L2?
CURR:DYN:L2? MAX
CURR:DYN:L2? MIN
CURRent:DYNamic:T1
Type: Channel-Specific
Description: Set duration parameter T1 for constant current dynamic mode.
Setting Syntax: CURRent:DYNamic:T1<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: CURR:DYN:T1 10ms Set the dynamic duration T1 = 10ms.
CURR:DYN:T1 2 Set the dynamic duration T1 = 2s.
CURR:DYN:T1 MAX Set the dynamic duration T1 as
maximum value.
CURR:DYN:T1 MIN Set the dynamic duration T1 as
minimum value.
Query Syntax: CURRent:DYNamic:T1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: CURR:DYN:T1?
CURR:DYN:T1? MAX
CURR:DYN:T1? MIN
CURRent:DYNamic:T2
Type: Channel-Specific
Description: Set duration parameter T2 for constant current dynamic mode.
Setting Syntax: CURRent:DYNamic:T2<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: CURR:DYN:T2 10ms Set the dynamic duration T2 = 10ms.
CURR:DYN:T2 2 Set the dynamic duration T2 = 2s.
CURR:DYN:T2 MAX Set the dynamic duration T2 as
maximum value.
CURR:DYN:T2 MIN Set the dynamic duration T2 as
minimum value.
Query Syntax: CURRent:DYNamic:T2?[<space><MAX | MIN>]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals