3-38 Setting Up the Software
Following are sample configuration settings for Virtua in the
DICOM Q/R SCP device:
IP Address = 192.168.1.3
Port = 104
AE Title = QR_JOB_PROFILE
Testing the Query/Retrieval Setup
Use the following steps to test the Query/Retrieval setup of both the
DICOM Q/R SCP device and Virtua.
1. Test the Virtua configuration on the DICOM Q/R SCP device by
initiating a study store from the DICOM Q/R SCP device,
specifying Virtua as the destination. Verify that the study is
received by Virtua and that it can be recorded to disc.
2. Go to the
Query/Retrieve
tab on the Virtua Studies screen to
make sure that the tab is available and that a default
configuration file is created.
3. Configure the DICOM Q/R SCP device and Virtua AE titles in the
qrscu.default.txt
file on the Virtua SmartDrive. For a list of the
parameters in this file, refer to Table 3-6 on page 3-35.
4. Test the Query functionality by performing a query. For
information about how to perform a query, refer to “Querying
and Retrieving Studies” on page 4-4.
•
If an error message displays, then the DICOM Q/R SCP
device information in the Virtua
qrscu.default.txt
file is
probably incorrect.
•
If no error message displays, then the query was successful,
and the list of matching studies is displayed.
•
If no results are returned, try another query with different
criteria. Consult with someone familiar with the DICOM Q/R
SCP device to understand how their parameters are
assigned, and what would make for a legitimate study query.