3-36 Setting Up the Software
destination_ae_title = The Destination AE Title for Virtua. This is the AE Title name
that will be used to designate the Job Profile for studies that are
retrieved.
This AE Title must be configured in the DICOM Q/R SCP
device as a valid Store destination.
It is recommended that the Destination AE Title be the same as
the Calling AE Title.
patient_name = The default entry for the Patient Name field in the Query form.
This parameter can be left blank.
patient_id = The default entry for the Patient ID field in the Query form.
This parameter can be left blank.
accession_number = The default entry for the Accession Number field in the Query
form.
This parameter can be left blank.
query_timeout_seconds = The number of seconds Virtua will wait for query results before
aborting the request.
move_timeout_seconds = The number of seconds Virtua will wait for retrieval results
before aborting the request.
max_returned_results = The maximum number of matching studies that will be
accepted by Virtua in response to a query.
query_level = The DICOM query level.
Currently, this parameter entry must be STUDY.
information_model = The DICOM information model.
Currently, this parameter entry must be STUDY ROOT.
Table 3-6. Query/Retrieve Profile Format and Parameters (Continued)
Parameter Description