Manufacturer’s Declaration
Operator’s Manual 101
Table11-10.Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
Condition AC Line Polarity Line Cord
Neutral
Line Cord IEC 60601-1
IEC 60601-1
ANSI/AAMI
ES 60601-1: 2005
Normal Normal Closed Closed 500 µA 300 µA
Single Fault Open Closed 1000 µA
Closed Open
Normal Reversed Closed Closed 500 µA 300 µA
Single Fault Open Closed 1000 µA
Closed Open
Enclosure Leakage Current
Condition AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
ANSI/AAMI ES 60601-1: 2005
Normal Normal Closed Closed 100 µA
Single Fault Open Closed 500 µA
Closed Open
Normal Reversed Closed Closed 100 µA
Single Fault Open Closed 500 µA
Closed Open