Operation
34 Operator’s Manual
Fast response mode — Responds to changes in blood oxygen saturation
within two (2) to four (4) seconds. This mode can be particularly helpful for
situations that require close monitoring.
Figure4-7.Response Mode Menu
Note:
When in the Fast Response Mode, the monitoring system may produce more SpO
2
and pulse rate alarms than expected.
Trend Data Download
Access this menu option to download patient trend data. Reference Trend
Data Download, p. 5-49.
Delete All Trend Data
Access this menu option to delete all patient trend data from memory.
To delete all trend data:
1. Access the OPTIONS Menu.
2. Rotate the jog dial to highlight DELETE ALL TREND DATA.