Clinical Studies
106 Operator’s Manual
A.6 Conclusion
The pooled results indicate that for a saturation range of 60-80% for SpO2, the
acceptance criterion was met for the monitoring system when tested with
MAX-A, MAX-N and MAX-FAST sensors. The pooled results indicate that for a
saturation range of 70-1 00% for SpO
2, the acceptance criterion was met.