Do not immerse or wet the pulse oximetry sensor.
Caution:
Nellcor™ adhesive pulse oximetry sensors are intended for single-patient use
only. Do not reuse pulse oximetry sensors.
When selecting a Nellcor™ pulse oximetry sensor, consider the following
items: patient’s weight and activity level, the adequacy of perfusion, and the
available sensor sites, the need for sterility, and the anticipated duration of
monitoring. Use the following table for selection or contact Covidien or a local
Covidien representative. Reference Sensor Performance Considerations, p. 6-
65. Use the Nellcor™ pulse oximetry interface cable to connect the pulse oxim-
etry sensor to the monitoring system.
Table9-1.Nellcor™ Pulse Oximetry Sensor Models and Patient Sizes
Nellcor™ Pulse Oximetry Sensor SKU Patient
Size
Nellcor™ Preemie SpO
2 Sensor, non-adhesive (Single-patient use)
SC-PR <1.5 kg
Nellcor™ Neonatal SpO
2 Sensor, non-adhesive (Single-patient use)
SC-NEO 1.5 to 5 kg
Nellcor™ Adult SpO
2 Sensor, non-adhesive (Single-patient use)
SC-A >40 kg
Nellcor™ Adult-Neonatal SpO
2 Sensor with Wraps (Reusable with adhesive)
OXI-A/N <3 or >40 kg
Nellcor™ Pediatric-Infant SpO
2 Sensor with Wraps (Reusable with adhesive)
OXI-P/I 3 to 40 kg
Nellcor™ Pediatric SpO
2 Sensor, Two Piece (Sterile, single-use only)
P 10 to 50 kg
Nellcor™ Neonatal-Adult SpO
2 Sensor, Two Piece (Sterile, single-use only)
N <3 or >40 kg
Nellcor™ Adult SpO
2 Sensor, Two Piece (Sterile, single-use only)
A > 30 kg
Nellcor™ Neonatal-Adult SpO
2 Sensor (Sterile, single-use only)
MAX-N <3 or >40 kg
Nellcor™ Infant SpO
2 Sensor (Sterile, single-use only)
MAX-I 3 to 20 kg
Nellcor™ Pediatric SpO
2 Sensor (Sterile, single-use only)
MAX-P 10 to 50 kg
Nellcor™ Adult SpO
2 Sensor (Sterile, single-use only)
MAX-A >30 kg
Nellcor™ Adult XL SpO
2 Sensor (Sterile, single-use only)
MAX-AL >30 kg