Factory Defaults
Operator’s Manual 43
4.5.2 Visual Alarm Indicators
Visual alarms appear on the screen in order of highest priority, regardless of
any audible alarm status. Reference Table4-2. on page 4-40.
4.6 Factory Defaults
The monitoring system ships with factory default settings. To set different insti-
tutional default settings, contact a qualified service technician.
Table4-4. Parameter Ranges and Factory Defaults
Parameter Ranges/Selection Factory Default
Adult Pediatric Neonatal
SpO
2
%SpO2 Upper Alarm Limit 21 to 100% (1% steps) 100% 95%
%SpO
2 Lower Alarm Limit 20 to 99% (1% steps) 90% 85%
%SpO
2 Limit Alarm Inhibition On, Off Off
SatSeconds™ Alarm Off, 10, 25, 50, 100 100
Pulse Rate
Pulse Rate Upper Alarm Limit 30 to 245 bpm (5 bpm steps) 170 bpm 200 bpm
Pulse Rate Lower Alarm Limit 25 to 240 bpm (5 bpm steps) 50 bpm 75 bpm 100 bpm
PR Limit Alarm Inhibition On, Off Off
Tabular Trends
Scroll 1, 5, 100, 500 1
Graphical Trends
SpO
2 On, Off On
PR On, Off On
Others
Patient Mode Adult, Pediatric, Neonatal Adult
Alarm Volume 1, 2, 3, 4, 5, 6, 7, 8 5
Key Beep Volume Off, 1, 2, 3, 4, 5, 6, 7 4
Pulse Volume Off, 1, 2, 3, 4, 5, 6, 7 4