CoCo-80 User Manual
99
Using Integration
Integration can be enabled in the Input Channel table. In CoCo, to set up the
built-in integration or double integration, you must set two engineering units. The
first one is for sensor sensitivity; the second for the engineering unit after the
integration or differentiation. For example, you can choose either g or m/s
2
as the
engineering unit used for the accelerometers. After the double integration, you
can choose one from the list of meter, cm, mm or other displacement units for
displacement presentation.
Example
First select the Acceleration or Velocity in the input channel table:
Figure 108. Sensor sensitivity without integration.
Then select the engineering unit of the sensor to be used and set its sensitivity.
Under the item of Selection integration or differentiation, select appropriate item.
The example shows Double Integration to Displacement is selected