CoCo-80 User Manual
40
Figure 43. Arbitrary waveform setup.
In the arbitrary waveform setup, the duration is fixed by the number of points in
the arbitrary data file and the sampling rate in use. The Quiet Zone is the time
with “zero” output between two arbitrary waveform pulses. The Peak Output
Level is the normalized maximum volt for the output waveform. Regardless of the
value in the arbitrary file, it is always normalized to this peak level volt.
Param -> Acquisition Mode is used to configure the trigger and how the data
blocks are captured from the conditioned time streams into the signal analyzer
phase. The selections are:
1. Free Run
2. Continuous after Trigger
3. Single Shot without Trigger
4. Single Shot with Trigger
5. Manual-arm Trigger
6. Auto-arm Trigger
It is important to note that the Acquisition Mode is designed for signal analysis
functions only, such as spectrum measurements. The data conditioning process is
not affected by the Acquisition Mode. For example data recording will continue
uninterrupted regardless of the Acquisition Mode. Acquisition Mode setting is
dependent on the selected CSA.
Param -> Schedule Setup is used to configure automated test schedules.