Cygnus DIVE Operating Manual M2-DIVE-M-ENG_Iss5.docx
Page 21 of 118
Twin Element Probes
Twin Element probes can be used to take thickness measurements
on both Coated and Un-Coated metals.
When measuring on Uncoated surfaces Single Echo mode can be
used. Single Echo mode will also work best when measuring Heavy
Corrosion – heavy corrosion will absorb and scatter the ultrasound
signal so there is often only enough ultrasound reflected back to
the probe to provide a single echo.
Identifying Single or Twin Element Probes
You can easily identify each type of probe by looking at the probe
face;
The Twin Element probe has two
visible sections and NO
protective membrane
The Single Element prob
e has a
single black face with a
protective membrane
Single
Crystal
Probe
Twin
Crystal
Probe