EasyManuals Logo

DEWESOFT IOLITE User Manual

DEWESOFT IOLITE
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
IOLITE®
TECHNICAL REFERENCE MANUAL
5.11.1. 3xMEMS-ACC: Specifications
MEMS
Accelerometer
3xMEMS-ACC-8g
3xMEMS-ACC-40g
Min.
Typ.
Max.
Min.
Typ.
Max.
Measurement
ranges
+-2 g
+-8 g
+-10 g
+-40 g
-3 dB bandwidth
1000 Hz
1000 Hz
Sample rate
4 kHz
4 kHz
Dynamic range
96 dB
96 dB
Noise density
(lowest range)
25 µg/√Hz
75 µg/√Hz
Residual noise
(+-2 g range, 50
Hz bandwidth)
100 µg RMS
Residual noise
(+-2 g range, 125
Hz bandwidth)
150 µg RMS
Offset error
-25 mg
10 mg
25 mg
-100 mg
30 mg
100 mg
Offset
temperature
drift (-20 .. 60
degC)
-0.15 mg / degC
+-0.02 mg / degC
0.15 mg / degC
-0.75 mg / degC
+-0.2 mg / degC
0.75 mg / degC
Semsitivity
temperature
drift (-20 .. 60
degC)
+-0.01 % / degC
+-0.01 % / degC
Linearity error
(smallest range)
0.1 % FS
0.1 % FS
Crossaxis
sensitivity
-1 %
1 %
1 %
MEMS
Inclinometer
Typ
Accuracy and
resolution valid
in range
+-15 deg
Accuracy and
resolution valid
for bandwidth
< 0.1 deg
Relative accuracy
(23 degC)
0.01 deg
Resolution
0.001 deg
IOLITE® V20-1 136/153

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT IOLITE and is the answer not in the manual?

DEWESOFT IOLITE Specifications

General IconGeneral
BrandDEWESOFT
ModelIOLITE
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals