ICS-90 Ion Chromatography System
3-2 Doc. 031851-05 12/04
2. Empty the entire bottle of concentrate into the reservoir.
3. Fill the concentrate bottle with DI water and empty the contents into the
reservoir.
4. Continue to pour DI water into the reservoir until it is filled all the way to the
top. If a few drops spill over, then it is full enough.
5. Insert the stopper assembly tubing into the reservoir and tighten the stopper.
For additional information, refer to the Installation Instructions and
Troubleshooting Guide for the Displacement Chemical Regeneration (DCR) Kit
(Document No. 031664).
3.3 Pressurizing the Eluent Reservoir
1. Before beginning operation, verify that the clean gas supply is regulated to
between 0.55 and 0.83 MPa (80 and 120 psi) and is connected to the gas inlet
of the ICS-90 air regulator (see Figure 3-1
).
2. Verify that the gas line from the eluent reservoir is connected to the ICS-90 air
regulator outlet line.
3. Make sure the eluent and regenerant reservoirs are filled, the caps are tightly
screwed onto the reservoirs, and the lines connecting the eluent and
regenerant reservoirs to the ICS-90 lines are securely connected.
4. Pull out the air regulator knob and turn it clockwise to pressurize the eluent
reservoir.
5. Adjust the pressure to between 30 and 40 kPa (5 and 6 psi). Push the regulator
knob back in.
Gießen Sie bei säurehaltigen Konzentraten (beispielsweise dem
Kationenelutionsmittel) das Konzentrat immer in entionisiertes
Wasser und nicht in den leeren Behälter.
The regenerant reservoir must be filled all the way to the top at all
times.
Do not shake the regenerant reservoir to mix the contents. The ICS-90
will stratify the contents.