EasyManua.ls Logo

Dionex ICS-90 - Peak Ghosting

Dionex ICS-90
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Troubleshooting
Doc. 031851-05 12/04 4-7
Plugged or damaged fitting
Isolate the faulty fitting by loosening fittings, one by one, until the pressure
returns to normal. Repair or replace the fitting (see Section 5.2
).
Flow rate through the columns too high
Measure the pump flow rate and set it to the correct rate (see Section B.14
).
Clogged column bed supports
Refer to the instructions in the column manual for troubleshooting guidance.
The column manual is included on the Dionex Reference Library CD-ROM
(P/N 055405), located in the ICS-90 Ship Kit.
Contaminated columns
Clean the columns as instructed in the column manual.
Plugged Rheodyne valve passages
Rebuild the injection valve (see Section 5.9
).
4.8 Peak “Ghosting”
Ghosting is the appearance of extraneous peaks in a chromatogram. These may be
late-eluting peaks from a previous injection or they may result from a
contaminated, malfunctioning, or incorrectly installed injection valve. These
peaks may co-elute with peaks of interest, resulting in nonreproducible peak
heights/areas.
Insufficient time between sample injections
Wait until the previous sample has been completely eluted before making
another injection.
Insufficient flush between samples
Flush the sample loop with at least 10 loop volumes of deionized water or
sample between sample injections.
Incorrect or contaminated standards
Remake standards.
Incorrect or contaminated eluent
Remake the eluent (see Section 3.1
).

Table of Contents

Other manuals for Dionex ICS-90

Related product manuals