ICS-90 Ion Chromatography System
5-20 Doc. 031851-05 12/04
10. Close the front door.
11. Reconnect the power cord and turn on the ICS-90 power.
12. Calibrate the cell (see Section 5.1.2
).
5.11 Installing a New Suppressor
Refer to the suppressor manual for guidance about when to replace a suppressor.
The suppressor manual is included on the Dionex Reference Library CD-ROM
(P/N 055405), located in the ICS-90 Ship Kit.
1. Turn off the pump from the ICS-90 Control panel in Chromeleon.
2. Open the front door of the ICS-90.
3. Disconnect the four eluent and regenerant lines from the suppressor.
4. Slide the suppressor to the left to detach it from the component mounting
panel.
5. Slide the new suppressor (AMMS III suppressor, P/N 056751; CMMS III
suppressor, P/N 056753) to the right until it locks into place on the mounting
panel.
6. Connect the four eluent and regenerant lines to the new suppressor.
7. Close the ICS-90 front door.
8. Prime the pump (see Section B.13
).