EasyManua.ls Logo

Dionex ICS-90 - Changing the Sample Loop

Dionex ICS-90
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5•Service
Doc. 031851-05 12/04 5-9
5.3 Changing the Sample Loop
Figure 5-3 shows the injection valve connections. The injection valve is plumbed
at the factory with all tubing and fittings for connection to the pump, injection
port, and column. A 10 µL PEEK sample loop (P/N 042857) is installed at the
factory between ports L and L. Always replace the pre-installed 10 µL loop with
another of the same size.
1. Turn off the pump from the ICS-90 Control panel in Chromeleon.
2. Open the ICS-90 front door.
3. Disconnect the sample loop from ports L and L on the injection valve (see
Figure 5-3
).
4. Install the new sample loop between ports L and L on the injection valve.
5. Make sure the loop is tightly coiled so the door can close securely.
6. Close the door.
7. Turn on the pump from the ICS-90 Control panel.
Figure 5-3. Injection Valve Plumbing

Table of Contents

Other manuals for Dionex ICS-90

Related product manuals