Dionex ICS-90 - Nonreproducible Peak Height or Retention Time; Abnormal Retention Time or Selectivity

Dionex ICS-90
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICS-90 Ion Chromatography System
4-8 Doc. 031851-05 12/04
Malfunctioning injection valve
Contact Dionex for assistance.
4.9 Nonreproducible Peak Height or Retention Time
Column overloading
Dilute the sample (see Section 3.4.3
).
Liquid leaks
Locate and eliminate the leaks (see Section 4.2
).
Incomplete or imprecise filling of the sample loop
1. Fill the sample loop until excess sample exits the waste line.
2. Inspect the 1 cc syringe (P/N 016388) and replace if damaged.
Pump not primed properly
Prime the pump (see Section B.13
).
4.10 Abnormal Retention Time or Selectivity
Contaminated or incorrect eluent
Remake the eluent, using concentrated eluent and ASTM filtered, Type I (18-
megohm) deionized water.
Contaminated or degraded sample
Take appropriate precautions when preparing and storing samples to prevent
contamination and degradation (see Section 3.4
).
Contaminated column
1. Clean the column as instructed in the column manual. The column
manual is included on the Dionex Reference Library CD-ROM
(P/N 055405), located in the ICS-90 Ship Kit.
2. If cleaning is unsuccessful, replace the column.