4 Troubleshooting
Doc. 031851-05 12/04 4-3
OVER CONDUCTIVITY
This message usually appears for the first few minutes after the ICS-90 is
started. If the message appears at other times, it indicates a problem.
To troubleshoot:
1. Perform the suppressor QuickStart procedure:
a. Disconnect the separator column outlet line and connect it to the
DS5 cell inlet line.
b. Turn on the pump from the ICS-90 Control panel in Chromeleon
and allow the suppressor to regenerate for 5 minutes.
c. Turn off the pump and allow the suppressor to hydrate for
5 minutes.
d. Reconnect the separator column outlet line and cell inlet line.
2. If the high background persists:
Prepare fresh eluent.
Prepare fresh regenerant.
Verify the cell calibration by pumping deionized water through
the cell (bypassing the column(s) and suppressor). The total
background should be less than 1.0
µS.
Perform the suppressor off-line regeneration procedure. For
instructions, refer to the suppressor manual.
If the problem persists, replace the suppressor (see Section 5.11).
DATA BUFFER OVERFLOW ERROR
Various electronics-related problems can cause the following error message to
be displayed in the Audit Trail:
If this warning appears, call Dionex for assistance. The ICS-90
electronics components cannot be serviced by the user.
Conductivity exceeds limit.
Module data buffer overflow. Data may have been lost.