ICS-90 Ion Chromatography System
5-22 Doc. 031851-05 12/04
3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated
3.15 amps (P/N 054745). Dionex recommends always replacing both fuses.
4. Reinsert the fuse holder into its compartment. The fuse holder is keyed to fit
only in its proper orientation. Apply sufficient pressure evenly against the
holder to engage the two locks. The holder is flush against the panel when
both locks are engaged.
5. Reconnect the main power cord and turn on the power.