ICS-900 Operator’s Manual
94 Doc. 065215-02 9/09
A.4 Front and Rear Panel LEDs
Front Panel LEDs
Rear Panel LED
A.5 Pump
Power Indicates when the power is on.
Ready Indicates when the ICS-900 is ready to acquire data; flashes if the
system check fails.
Run Indicates when the ICS-900 is running/acquiring data; flashes if an
error occurs.
Link Indicates when there is communication between the ICS-900 and
Chromeleon or Chromeleon Xpress; flashes when data is being
transmitted or received.
Type Serial dual-reciprocating piston, microprocessor-controlled
constant stroke, variable speed
Construction Chemically inert, metal-free PEEK pump heads and flow path.
Compatible with aqueous eluents of pH 0 to 14 and reversed-phase
solvents
Operating
Pressure
0 to 35 MPa (0 to 5000 psi)
Flow Rate
Range
0.00 to 5.00 mL/min, in 0.01 mL/min increments (without
changing pump heads)
Flow Precision <0.1%
Flow Accuracy <0.1%
Pressure Ripple <1%
Eluent Bottle
Pressure
None required