5•Service
Doc. 065215-02 9/09 67
3. If the controls on the Wellness panel are disabled, select Connect to
timebase on the Control menu and select the ICS-900 timebase.
5.1.2 Wellness Panel Features
The Wellness panel provides the following features:
System Status The check box indicates whether the ICS-900 is connected to
Chromeleon or Chromeleon Xpress. When there is no
connection, the check box is clear and the remaining controls
on the Wellness panel are disabled.
Click the Pump button to turn the pump on and off.
Click the Log Serial # button to record the ICS-900 serial
number in the Audit Trail.
The Calibration or Diagnostic box is green while a
calibration or diagnostic procedure is running.
The current total system Pressure, pump Flow rate, and
electrochemical detector (ECD) conductivity are displayed.
Audit Trail
The Audit Trail displays an account of every event that occurs
during ICS-900 operation. This includes errors, status
messages, operational events, etc.
Reset
Calibration
Parameters
Click the View/Reset Cal Values button to open a window
that shows the current calibration values, the previous values,
and the factory-set values. You can reset the current
calibration values to the previous or factory-set values from
the window.