Doc. 065215-02 9/09 65
5•Service
This chapter describes ICS-900 Ion Chromatography System (ICS-900) service
and repair procedures that users may perform. Procedures not included here,
including electronics-related repair procedures, must be performed by Dionex
personnel. For assistance, contact Dionex Technical Support. In the U.S., call 1-
800-346-6390. Outside the U.S., call the nearest Dionex office.
Before replacing any part, refer to the troubleshooting information in Chapter 4
to
correctly identify the cause of the problem.
5.1 Diagnostics and Calibrations
The Wellness panel (see Figure 5-2) provides controls for performing diagnostic
and calibration functions. This section provides an overview of Wellness panel
features. For calibration procedures, see the following sections:
• “Calibrating the Pressure Transducer” (Section 5.1.3)
• “Calibrating the Cell” (Section 5.1.4)
• “Calibrating the Flow Rate” (see Section 5.1.5)
5.1.1 Opening the Wellness Panel
1. In the Chromeleon Browser, expand the Dionex
Templates\Panels\Wellness folder.
Substituting non-Dionex parts may impair system performance,
thereby voiding the product warranty. Refer to the warranty statement
in the Dionex Terms and Conditions for more information.