EasyManua.ls Logo

Dionex ICS-900 - Cleaning Eluent Bottles; Changing the Sample Loop

Dionex ICS-900
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICS-900 Operator’s Manual
76 Doc. 065215-02 9/09
5.4 Cleaning Eluent Bottles
Before preparing new eluent, all eluent bottles should be rinsed thoroughly (inside
and out) with ASTM Type I (18 megohm-cm) filtered and deionized water. If a
bottle still appears dirty, or has a slimy film on the inside, clean it as instructed
below.
1. Dispose of any remaining chemicals according to municipal regulations.
2. Rinse the bottle (inside and out) with ASTM Type I (18 megohm-cm) filtered
and deionized water.
3. Rinse the inside of the bottle with isopropyl alcohol or methanol.
4. If algae or bacteria have left a slimy film on the bottle, use an algicide or
disinfectant (dilute hydrogen peroxide, etc.).
5. Rinse cleaning chemicals out of the bottle with ASTM Type I (18 megohm-
cm) filtered and deionized water.
6. Dry the bottle with clean, particulate-free air.
5.5 Changing the Sample Loop
A 10 μL PEEK sample loop (P/N 042949) is installed on the injection valve at the
factory. Unless you are using a 2 mm column, replace the sample loop with
another 10 μL PEEK sample loop. If you are using a 2 mm column, refer to the
column manual for the size of loop required. Column manuals are included on the
Dionex Reference Library CD-ROM (P/N 053891)
1. Turn off the pump from
the ICS-900 Control
panel.
2. Open the ICS-900 front
door.
3. Disconnect the sample
loop from ports
L and L on
the injection valve (see
Figure 5-6
).
Figure 5-6. ICS-900 Injection Valve
Sample Loop

Table of Contents

Related product manuals