ICS-900 Operator’s Manual
32 Doc. 065215-02 9/09
After the analysis begins, do not mix the contents of the regenerant bottle.
10.Verify that the liquid lines from the ELUENT and REGEN bottles are
connected to the corresponding lines from the ICS-900 (see Figure 3-4
).
3.6 Priming the Pump
Prime the pump if the eluent has been changed, the eluent line contains air, or the
pump heads are dry (for example, after servicing).
NOTE If the eluent line is empty or the pump heads are
completely dry, you can use a 10 cc syringe (P/N 054578)
to prime the pump (see page 34
).
1. Verify that the ELUENT and REGEN bottles are filled, the bottle caps are
installed and hand-tightened, and the liquid lines are connected to the bottles.
2. Verify that the waste lines are directed to a waste container.
Figure 3-4. Eluent and Regenerant Bottle Liquid Line Connections
ELUENT Bottle Out
REGEN Bottle Out
REGEN
Bottle In