5•Service
Doc. 065215-02 9/09 91
11. Reconnect the power cord and turn on the ICS-900 power.
12. Calibrate the cell (see Section 5.1.4
).
5.12 Replacing the Suppressor
Refer to the suppressor manual for guidance about when to replace a suppressor.
Suppressor manuals are included on the Dionex Reference Library CD-ROM
(P/N 053891).
1. Turn off the pump.
2. Open the front door of the ICS-900.
3. Disconnect the four eluent and regenerant lines from the suppressor.
4. Slide the suppressor to the left to detach it from the component mounting
panel.
5. Slide the new suppressor to the right until it locks into place on the mounting
panel.
6. Connect the four eluent and regenerant lines to the new suppressor.
7. Close the ICS-900 front door.
8. Prime the pump (see Section 3.6
).